Can SONICC quantify crystalline drug in drug–polymer powders and solid dispersions, and does its signal scale linearly with crystallinity?
Yes for naproxen–HPMCAS physical mixtures: the signal followed crystalline content linearly (slope 0.98, R² 0.998), with accuracy close to PXRD and Raman. In a 25% naproxen dispersion it agreed with both over about 24–54% crystallinity, but linearity there was not tested by regression. ···
The authors conclude that SONICC "has been found to enable sensitive quantification" of crystalline naproxen in HPMCAS, in cryomilled physical mixtures and in a solvent-evaporated 75/25 dispersion, because amorphous drug and polymer give almost no SHG background and crystals stay visible in a blend that is 99.9% polymer. ··· The fresh dispersion "appears to be amorphous" by PXRD and Raman yet held crystals by SONICC, a miss the authors "most likely" attribute to dilution of the drug by the polymer, and during storage at 60 °C SONICC followed crystallization in "good agreement" with both methods. ··· From these comparisons "it is clear" to the authors that SHG scales linearly with crystallinity in these dispersions and "reasonable to expect" this to be the general trend; they see SONICC as "well suited" to dispersions of chiral compounds but "arguably best used as a complement" to conventional methods. ··· Three links were not measured: linear scaling within the dispersion itself, tested by regression only on physical mixtures; the crystalline content of the fresh dispersion by an independent method; and the crystal perfection and size on which the incoherent reading of linear scaling and the submicron-domain claim rest. ···
01Why this study
Even low levels of crystals can seed crystallization in storage and dissolution and cancel the amorphous advantage.
Rodriguez-Hornedo 1999, ref 6; refs 7, 8
Most solid-state techniques cannot detect crystallinity below about 1%; in dispersions the polymer also dilutes the drug and can interfere.
Taylor 1998, ref 14; refs 2, 15–17 ·
Amorphous material gives no net SHG. SHG had measured crystallinity in PVP blends, and the group's SONICC found crystals down to 0.04% in cryomilled griseofulvin.
Strachan 2004, ref 21; Wanapun 2011, ref 25 ··
SHG can scale between linearly and quadratically with crystallinity; linear scaling in powders rested on a few PXRD comparisons.
Wanapun 2010, ref 24 ·
More detail
What was already known. The introduction and theory section build on these points:
- Amorphous forms help but crystallize. Amorphous solids are increasingly studied to raise the apparent solubility, dissolution and potential bioavailability of poorly soluble drugs, but their development is held back by physical instability and by a lack of tools that detect low levels of crystallinity. (Hancock 1997, ref 3; Shan 2008, ref 4; Hancock 2002, ref 5)
- Trace crystals seed further crystallization. Even low levels of crystals can act as nucleation centres that speed crystallization during storage and dissolution and so cancel the advantage of the amorphous form. (Rodriguez-Hornedo 1999, ref 6; Fletcher 1963, ref 7; Vekilov 2010, ref 8)
- Polymers stabilise the drug but hide its crystals. Adding a polymeric crystallization inhibitor to form a solid dispersion is a common approach (refs 9–13). In a dispersion, early crystallization is hard to detect because the polymer dilutes the crystallizing drug (little crystallizing material per volume) and can interfere with the analytical technique.
- Common methods stop near 1%. DSC, PXRD, microscopy, Raman, infrared and solid-state NMR are used together, but most cannot detect crystallinity below about 1%. (Byrn 1999, ref 2; Taylor 1998, ref 14; refs 15–17) Solid-state NMR is sensitive but takes hours, during which crystallinity can change; DSC heating can change the sample; TEM and SEM need sample preparation, are hard to use on powders and probe small fields that may miss trace crystals. The authors therefore look for a quick, reliable method that needs little sample preparation and gives both microscopic and macroscopic information on powders. (refs 2, 14–17)
- SHG is selective for crystals. SHG turns light into light of twice the energy, but only in material that lacks inversion symmetry: surfaces and most space groups of chiral crystals. In an isotropic sample the contributions of randomly oriented molecules cancel, so amorphous materials, liquids, solutions and centrosymmetric crystals give no net coherent SHG. SHG had been used for protein screening and chiral crystallization (refs 18–20).
- Earlier SHG work on pharmaceutical solids. Rades and coworkers used SHG to quantify polymorphs with a calculated LOD below 5% and to quantify crystallinity in PVP blends with enalapril maleate and ranitidine hydrochloride; for PVP–enalapril maleate form II the ensemble LOD was 0.12%. (Strachan 2004, ref 21; Strachan 2005, ref 22; Rawle 2006, ref 23) ·
- The group's earlier SONICC work. SONICC detected crystallites of 90 nm in an optically transparent medium (Wanapun 2010, ref 24) . In cryomilled griseofulvin powder it followed the loss of crystallinity in agreement with PXRD and detected crystals as low as 0.04%, well below the PXRD limit (Wanapun 2011, ref 25) .
- Dispersions were untested. Published work covered pure drugs or drug–polymer blends; SONICC had not been used on solid dispersions. (refs 21–25)
- Linear scaling was not expected. In pure powders, linear scaling (integrated SHG proportional to crystallinity) was suggested by PXRD comparisons at a few crystallinity values. In ordered thin films and pristine single crystals SHG scales quadratically, so linear scaling would imply incoherent combination of SHG within microcrystals. Earlier measurements used unfocused or gently focused light, and it is "not clear" that linearity holds when the focal volume (about 1 µm) is of the order of the crystal size. (Wanapun 2010, ref 24)
- Theory: between linear and quadratic. Two sources in phase add their fields, so intensity rises with the square of the number of sources; sources with random phases add their intensities, so the signal rises linearly ·. For epi-detected SHG the coherence length along the beam is about 100 nm (Wanapun 2010, ref 24) ; pristine growth beyond it makes intensity oscillate with thickness, and the overall scaling goes with volume to the 4/3 . Lattice defects over more than 100 nm, or polycrystals, turn each domain into an independent source and give linear scaling. The scaling therefore "may depend sensitively" on the crystals formed.
What the paper set out to do. Use SONICC to detect and quantify SHG from microcrystals in naproxen–HPMCAS physical mixtures and in a naproxen–HPMCAS solid dispersion, with PXRD and Raman as reference methods of known linear scaling, and so test experimentally how SHG scales with crystallinity in solid dispersions and how low it can detect. ··
02How they approached it
One drug and one polymer. SHG is calibrated on weighed crystalline naproxen–HPMCAS blends side by side with PXRD and Raman, the reference methods, and the three then follow one dispersion crystallizing at 60 °C. ···
What was done
Cryomill crystalline naproxen with HPMCAS at 0–100%, triplicate→sieve below 105 µm.·
Rotary-evaporate the 75/25 HPMCAS–naproxen dispersion→cryomill and sieve.
Image each blend by epi-detected SHG→integrated intensity normalised to a crystalline standard.·
PXRD patterns and Raman spectra of the same blends→OPLS calibrations.
Three-level prediction set read by all three methods.·
3D SHG imaging of a 0.1% blend→calculated detection limit.·
Spike a partially crystallized dispersion→SONICC recovery.
Fresh dispersion vs 75/25 physical mixture→store the dispersion at 60 °C for 48 h→SONICC, PXRD and Raman at each withdrawal.··
Crystallinity–time curves of the three methods: SONICC against a crystalline blend, PXRD and Raman against dispersion standards (Fig. 7).··
More detail
What was done
- Physical-mixture standards. HPMCAS and crystalline naproxen were weighed and cryomilled together (SPEX 6750 freezer mill, 8 Hz, 5 min at liquid-nitrogen temperature, cycles of 2 min milling and 1 min cooling; number of cycles not printed), warmed to ambient, handled in a dry-nitrogen glove box and sieved to below 105 µm. Levels ran from 0 to 100% crystalline naproxen in triplicate (read from Fig. 3: several below 2%, then 2, 5, 7.5, 15, 25, 35, 50, 65, 75, 90 and 100%). An independent prediction set at 3.5, 18 and 83% was milled the same way. ··
- Solid dispersion. 75 wt% HPMCAS and 25 wt% naproxen were dissolved in dichloromethane:ethanol 1:1 by weight, rotary-evaporated, dried under vacuum overnight, cryomilled and sieved to below 105 µm. Powder was stored at 60 °C and withdrawn at 0, 1, 2, 4, 6, 8, 12, 24 and 48 h; storage RH and container are not printed. ··
- Dispersion standards. For PXRD and Raman of the dispersion, crystalline naproxen was cryomilled with the dispersion to give 75/25 blends with different crystalline:amorphous naproxen ratios (levels not printed). SONICC of the dispersion was instead normalised to a 75/25 physical mixture of HPMCAS and crystalline naproxen. ·
- SONICC imaging. A custom-built SHG microscope: Spectra-Physics Mai Tai laser (800 nm, 100 fs, 80 MHz, 20 mW average power), 10× 0.3 NA objective, resonant mirror (about 8 kHz) and galvanometer scanning, one frame in 25 s with 1000 averages per line, 20 µm z-steps, resolution 4 × 4 × 20 µm³; SHG at 400 nm counted by photomultipliers. About 1–2 mg powder was held between slide and cover glass with 150 µm spacers, three fields of view per sample, and backward (epi) SHG was used for quantification. The low power was chosen "to ensure a linear dynamic range". ·
- SHG crystallinity. Average SHG intensity integrated over the field of view, minus the near-zero background of pure HPMCAS, divided by that of 100% crystalline naproxen (blends) or of the 75/25 crystalline blend (dispersion). ·
- 3D imaging. Z-stacks of 100% crystalline naproxen and of 0.1% crystalline naproxen in 99.9% HPMCAS, 3 min per 3D image. ·
- Detection limit. Calculated as the single-voxel signal at a signal-to-background of 3 on a sample with no naproxen, divided by the signal of 100% crystalline powder. ·
- PXRD. Shimadzu XRD-6000, Cu-Kα, 40 kV, 5–35° 2θ at 4°/min, step 0.04°, silicon external standard; full-pattern OPLS models (SIMCA P+ 12, mean-centred), one on the blends and one on the dispersion standards. ·
- Raman. Kaiser RamanRxn1 at 785 nm with a non-contact probe, 6–10 regions of the powder bed averaged, triplicate; OPLS on 3120–500 cm⁻¹ with standard normal variate correction and log transform, again one model on the blends and one on the dispersion standards. ·
- Spike test. A partially crystallized dispersion (storage history not printed) was spiked with a known amount of crystalline naproxen (amount not printed) and measured by SONICC. ·
Key equations
In phase, the fields of two sources add before squaring, so intensity grows with the square of the number of sources; with random phases the intensities add and the signal grows linearly. Both equations print source a twice where the text means sources a and b; the pair frames the scaling question that the calibrations test.··
A pristine crystal that grows beyond the 100 nm coherence length gains no signal from extra thickness, so its SHG grows with its area squared, that is with volume to the 4/3. Defects or polycrystals over more than 100 nm break this and give linear scaling instead.·
Relative crystallinity is the field-averaged SHG intensity minus the pure-HPMCAS background, divided by the same for the reference: 100% crystalline naproxen for the blends, a 75/25 crystalline blend for the dispersion. It holds only if the signal scales linearly, crystals are randomly oriented and the standards match the samples in particle size.···
The single-voxel signal at a signal-to-background ratio of 3 on a sample without naproxen, divided by the signal of 100% crystalline powder. This gives the 10 ppm figure, a calculation rather than an imaged sample.··
Terms used on this page
- SHG (second harmonic generation): conversion of laser light (here 800 nm) into light of twice the energy (400 nm) by material that lacks inversion symmetry.
- SONICC: second-order nonlinear optical imaging of chiral crystals, a laser-scanning SHG microscope that gives both images and an integrated intensity.
- Epi-detected: SHG collected backward through the focusing objective.
- Noncentrosymmetric crystal: a crystal without an inversion centre. Most chiral compounds crystallize this way; achiral compounds routinely do not.
- Coherent and incoherent SHG: sources in phase add their fields, so intensity grows with the square of their number; sources with random phases add their intensities, so the signal grows linearly.
- Coherence length: about 100 nm along the beam for epi-detected SHG; beyond it a thicker crystal adds no signal.
- OPLS: orthogonal projections to latent structures, a multivariate regression of whole PXRD patterns or Raman spectra on known crystallinity.
- RMSEE: root mean square error of estimation, the fit error on the calibration samples themselves, not a prediction error.
- Prediction set: samples left out of the calibration and used to test it. 95% prediction band: the range in which a new measurement is expected to fall.
- LOD: limit of detection.
- Physical mixture (PM) and solid dispersion (SD): crystalline drug blended with polymer, versus drug mixed into the polymer at the molecular level by solvent evaporation.
- Two crystallinity bases: for the blends, % crystalline naproxen in the whole powder; for the dispersion, % of the drug that is crystalline.
03Main findings
The SHG signal scales linearly with crystalline naproxen in the blends
In cryomilled naproxen–HPMCAS mixtures, SHG crystallinity followed the weighed content over 0–100% (slope 0.98, R² 0.998, RMSEE 2.68) and over 0–25% (slope 0.941, R² 0.996, RMSEE 1.90). ·
More detail
Calibration points lay close to the identity line at every level, for example about 3.5 and 5 at 5%, 13–15 at 15%, 44–52 at 50% and 88–97 at 90% (read from Fig. 3). The 95% prediction bands were about ±4.5 units over 0–100% and about ±2 units at 25% in the 0–25% insert, and the scatter grew with naproxen content. · Table 1 gives slope 0.98, R² 0.998 and RMSEE 2.68 for 0–100%, and slope 0.941, R² 0.996 and RMSEE 1.90 for 0–25%; the text gives slope 0.92 and R² 0.99 for 0–25%. · The abstract quotes R² 0.99 and a "root mean square error of prediction" of 2.7% over 0.1–100 wt%, which appears to be the rounded RMSEE. The authors read this as SHG intensity varying linearly with naproxen content, and HPMCAS "does not appear to interfere" with the signal. "Actual crystallinity" is the weighed composition; it was not checked after cryomilling.
Its accuracy matches PXRD and Raman
OPLS models on the same blends gave RMSEE 4.10 (PXRD) and 2.22 (Raman). On a test set at 3.5, 18 and 83%, SONICC read 3.86, 18.95 and 86.53%, errors of about 10% at the lowest level and about 5% above. ··
More detail
PXRD: slope 0.985, R² 0.981 (Q² 0.97), RMSEE 4.10, with prediction bands of about ±8 units, wider than for SONICC. Raman: slope 0.996, R² 0.998 (Q² 0.99), RMSEE 2.22, with bands of about ±2.5 units, narrower than for SONICC. ··· Prediction set (Table 2), actual → SHG / PXRD / Raman: 3.5 → 3.86 / "–" as printed / 3.74; 18 → 18.95 / 16.70 / 19.92; 83 → 86.53 / 85.73 / 88.75. The authors call the agreement good, give the SHG error as about 10% at the lowest level and 5% or less at the others, "comparable" with the other techniques, and note "undoubtedly errors" in sample preparation. The Discussion cites Table 1 for these values, which are in Table 2.
SONICC sees crystals at 0.1%; PXRD sees none below 5%
A blend of 0.1% crystalline naproxen in 99.9% HPMCAS showed about six discrete SHG-active crystals, and the calculated mean LOD was 10 ppm. PXRD showed no naproxen peaks below 5 wt%. ··
More detail
In 3D images, 100% crystalline naproxen gave SHG over the whole field, while the 0.1% blend gave about six discrete bright regions (count read from Fig. 4). The depth profile of one crystallite was a single peak of about 100 × 10³ counts/s at about 0 µm, with a base width of about 20–25 µm and nothing elsewhere between about −120 and +120 µm. The authors take this as "clearly indicating" that HPMCAS does not interfere and that "miniscule levels" of crystallinity are "readily detected". The LOD, from the single-voxel signal at signal-to-background 3 on a sample without naproxen divided by the 100% signal, is 1 × 10⁻³% crystallinity (10 ppm). · The authors place it well below common solid-state methods: laboratory PXRD is cited at about 5–7% (Nunes 2005, ref 30), Raman at as low as 1% for pure drug (Taylor 1998, ref 14), and the ensemble SHG LOD of Rades and coworkers at 0.12% (Strachan 2004, ref 21). ·· It is higher than the parts-per-trillion level reached in amorphous griseofulvin films because the polymer gives a nonzero background. · The intensities behind the LOD and the HPMCAS background value are not printed. In the PXRD blend calibration, samples below 5 wt% naproxen showed no detectable crystalline peaks, so a full-pattern OPLS fit was used.
A dispersion that looks amorphous already holds crystals
The fresh 75/25 HPMCAS–naproxen dispersion gave only a PXRD halo and no 863 cm⁻¹ Raman peak, yet SONICC showed sparse crystals. ··
More detail
The physical mixture at the same ratio gave sharp PXRD peaks (strongest at about 19.0° 2θ), the crystalline Raman peak at 863 cm⁻¹, and bright SHG over the whole field. The dispersion gave a broad halo peaking around 19–21° 2θ, a broad Raman band around 856 cm⁻¹, and sparse discrete SHG spots on a dark field. ·· By PXRD and Raman the dispersion "appears to be amorphous", while the SONICC images "clearly reveal" a "significant fraction of crystalline naproxen"; heavy dilution of the drug by the polymer "most likely" explains why PXRD and Raman miss it. · The signal is attributed to naproxen because the same material later developed naproxen PXRD peaks, as in the cryomilled griseofulvin study. · Normalised to the 75/25 crystalline blend, the SONICC reading at 0 h was about 24% of the drug ; the extractor calculates that this is about 6 wt% crystalline naproxen in the powder, close to the 5 wt% level below which the PXRD blends showed no peaks. The crystalline content was not checked by an independent method.
Added crystals add up in a spiked dispersion
A partially crystallized dispersion read 22.2% by SONICC; after spiking with crystalline naproxen it read 39.1%, against 41.1% expected.
More detail
The authors call this a recovery of "approximately 93%" (later "greater than 90%") and take it as showing the "linear additivity" of initial and added crystallinity, which "helps to further validate" SONICC for dispersions. · The 93% cannot be reproduced from the printed values: 39.1/41.1 is 95.1%, and recovery of the added part, (39.1 − 22.2)/(41.1 − 22.2), is 89.4%. The amount added and the storage history of the sample are not printed. ·
During storage, SONICC reads crystallinity first, then tracks PXRD and Raman
At 60 °C SONICC gave readings from 0 h (about 24%), while PXRD and Raman could not quantify before 2 h. By 48 h SONICC read about 48%, PXRD about 54% and Raman about 44%; crystallization was still incomplete. ··
More detail
PXRD showed no peaks before 2 h, a weak peak at about 19° 2θ at 2 h, and peaks at about 19, 22.5 and 23.7° 2θ by 12 h, all weaker than in the physical mixture. Raman showed a single broad band at about 856 cm⁻¹ at 0 h, and a shoulder at 863 cm⁻¹ from 2 h that grew into a peak. SONICC images showed sparse bright points at 0 h and more and larger bright regions at 2 h. With RMSEE of 5% (PXRD) and 7.4% (Raman) for the dispersion models, the authors judge that neither could estimate crystallinity before 2 h. · Time course (read from Fig. 7): SONICC about 24, 30, 31, 32.5, 34, 36, 38.5, 42.5 and 48% at 0, 1, 2, 4, 6, 8, 12, 24 and 48 h; PXRD about 27, 34, 37, 40, 43, 48 and 54% and Raman about 26, 27, 32, 33, 35, 39 and 44% from 2 to 48 h. The authors describe the kinetics as in "good agreement" (results, abstract) or as matching "reasonably well" (discussion), with SONICC falling between the other two and a discrepancy of this size being "common" (refs 16, 34, 35). SONICC lies between them from 4 h; at 2 h it is above both. At 75 wt% polymer, crystallization was incomplete after 48 h at 60 °C.
04Further observations
- PXRD's 95% prediction bands (about ±8) were wider than SONICC's (about ±4.5) and Raman's (about ±2.5). ··
- Dispersion-based models were less precise than the blend models: RMSEE 5% for PXRD and 7.4% for Raman. ·
- Raman marks crystalline naproxen by a peak at 863 cm⁻¹ against a broad amorphous band at 856 cm⁻¹. ·
- A 3D SONICC image took 3 min.
More detail
- Calibration scatter. Read from Figs. 3 and 5, the 95% prediction bands were about ±4.5 units for SONICC over 0–100% (about ±2 at 25% in the insert), about ±8 for PXRD and about ±2.5 for Raman. The authors place SONICC between PXRD (larger error) and Raman (smaller error). ··· PXRD over-read at the top of the range (about 101 and 108 at 100%).
- Dispersion models. The PXRD and Raman models built on dispersion standards had RMSEE 5% and 7.4%, against 4.10 and 2.22 for the blend models. Their levels, slopes and R² are not printed. ·
- Spectral markers. In the Raman 800–1000 cm⁻¹ region, crystalline naproxen shows a peak at 863 cm⁻¹ and the amorphous dispersion a broad band at about 856 cm⁻¹; a strong band at about 960 cm⁻¹ appears in both. During storage a lower band moved from about 830 cm⁻¹ (0 h) to about 825 cm⁻¹ (12–24 h), against about 823 cm⁻¹ in the physical mixture (read from Fig. 6). · The strongest naproxen PXRD peak lies at about 19.0° 2θ, the first to appear during storage. ·
- Speed. A frame takes from 10 s (about 200 averages per line) to several minutes (several thousand averages per line); a 3D image here took 3 min. The authors list data collection in seconds to minutes, no sample preparation and nondestructive measurement as advantages. ·
- Voxel intensities vary. The SHG intensity of a single voxel "can have a broad distribution" because crystal size and orientation vary.
05How the authors connect the findings
More detail
- SHG arises only where inversion symmetry is missing. Because the net signal is a coherent sum over all molecules, an isotropic sample cancels it: amorphous materials, liquids, solutions and centrosymmetric crystals give no net coherent SHG. Amorphous HPMCAS and amorphous naproxen therefore add almost no background, and chiral crystals are detected with high sensitivity.
- This explains the 0.1% result: discrete crystals show although 99.9% of the powder is polymer. In the dispersion, the polymer dilutes the crystallizing drug and can interfere with PXRD and Raman; the dispersion halo is "not necessarily" the sum of the pure amorphous halos, so PXRD "often requires" dispersion-based standards, and Raman spectra of dispersions are "generally not simple" to interpret and need chemometrics. Heavy dilution "most likely" explains why both missed the crystals in the fresh dispersion. ···
- SHG need not scale linearly with crystallinity. ·
Coherent sources. pristine lattice → fields of sources in phase add → intensity grows with the square of source number density, and pristine crystals growing beyond the 100 nm coherence length scale with volume to the 4/3 → a calibration that curves. ··
Incoherent sources. lattice defects over more than 100 nm or polycrystals → each domain acts as an independent source with random phase → intensities add → a linear calibration. ·
- The authors had noted that linear scaling would imply incoherent combination within microcrystals. From the comparisons with PXRD and Raman, "it is clear" that SHG scales linearly over the measured range in these dispersions, in line with earlier powder work (ref 24); the conclusions call this "helping to resolve a fundamental scaling ambiguity", and "it is reasonable to expect" linear scaling to be the general trend in drug crystallization studies. ·
- Because the response is linear and amorphous material is silent, SONICC can be calibrated with pure crystalline drug or a crystalline drug–polymer blend, without chemometrics or amorphous–crystalline standards.
What limits the method. The polymer background is near zero but not zero, which keeps the LOD (10 ppm) above that reached in pure amorphous griseofulvin films (parts per trillion). Nuclei below about 100 nm are "unlikely" to be detected. Readings can differ when crystallites in the sample differ in size from those in the standard, since larger domains give more SHG up to the coherence length (ref 36), and the assumption of random crystal orientation "might not" hold for very sparse early crystals. ·· Only SHG-active (mostly chiral) crystals are seen, no chemical or structural information is given, and crystalline chiral excipients can interfere.
Conclusion drawn. SONICC "has been found to enable sensitive quantification" of crystalline material in drug–polymer blends and solid dispersions, gives microscopic and macroscopic information, and detects early crystallization when the crystalline fraction is low and the domains are submicron. It is "well suited" to dispersions of chiral compounds, but "arguably best used as a complement" to conventional analyses rather than a replacement. ··
06What it means for formulation
- SONICC "would be especially advantageous" for ASDs at low drug loading, to choose polymer and process without long storage studies.
- Calibration standards must match the particle size of the samples.
- "Amorphous by PXRD and Raman" does not mean free of crystals, and the gap grows as drug loading falls.
More detail
- Trace crystals can seed crystallization in storage and dissolution, so detecting them early matters.
- SONICC would be "especially advantageous" for monitoring minute crystallinity in ASDs at low drug loading, where optimizing polymer and processing may otherwise need long storage before less sensitive methods detect crystals.
- Crystallinity can be quantified against pure crystalline drug or a crystalline drug–polymer blend, without amorphous–crystalline standards or complex chemometrics.
- Calibration standards should match the particle size range of the samples, as for Raman.
- The kinetics results show "potential utility" as an online monitoring tool.
- SONICC works only for drugs that crystallize in SHG-active space groups (most chiral compounds, routinely not achiral ones), and crystalline chiral excipients can interfere; it is "arguably best used as a complement" to PXRD, Raman and other methods. ·
- A dispersion that passes PXRD and Raman as amorphous can still carry crystals: here SONICC read about a quarter of the drug as crystalline at 0 h, about 6 wt% of the powder. At a lower drug loading the same fraction of the drug would sit further below the PXRD limit, so the risk of missing seeds grows as loading falls. ·
- Use SONICC numbers on a stored dispersion as a trend rather than an absolute value: they rest on a physical-mixture standard, and crystal size and orientation in the dispersion may differ from those in the standard. ··
- Before relying on SONICC for a given drug, confirm that its crystal form is SHG-active; a centrosymmetric form, or a crystalline chiral excipient, would mislead the reading.
07Open questions
- How much do crystallite size and orientation in a dispersion bias the reading against the standard? ·
- Linearity in the dispersion was not tested by regression, and its first reading was about 24% of the drug, not a trace. ·
More detail
- Differences in crystallite size between the dispersion and the standard can change SONICC values, and the same affects PXRD and Raman.
- The random-orientation assumption "might not" hold for very sparse crystals early in crystallization.
- Nuclei below about 100 nm are unlikely to be detected; SEM and TEM may see smaller crystals at relatively high number densities. ·
- With the current set-up, SONICC gives no chemical or structural information beyond percent crystallinity.
- Linear scaling was shown on physical mixtures; in the dispersion there is only agreement over about 24–54% crystallinity, with no regression against PXRD or Raman.
- Earlier detection in the dispersion was shown at about 24% of the drug; the 0.1% and 10 ppm figures come from physical mixtures, and 10 ppm was calculated, not imaged. ·
- SONICC on the dispersion was never run against the dispersion-based standards used for PXRD and Raman, so the claimed advantage of crystalline standards is untested.
- Not measured: SHG linearity with laser power at the measured crystallinities, SHG response to particle size, crystallite size in the dispersion and standards, and an online measurement. ···
- Where the crystals in the fresh dispersion came from (incomplete amorphization, or cryomilling after drying) was not examined. ·
- One drug, one polymer, one drug loading (25%) and one temperature were tested, and the storage RH is not printed. ·
08How solid
AI assessment, not a statement made in the paperFor the physical mixtures the case is strong: three methods calibrated on the same triplicate samples agree, and a test set and a spike test were run. For the dispersion, linearity and earlier detection rest on one time course without regression, starting at about 24% crystallinity rather than a trace. ···
More detail
- Linear SHG calibration on triplicate blends over 0–100%, with prediction bands and Table 1 statistics. ·
- PXRD and Raman models on the same samples give comparable accuracy, and the test set is predicted within about 5–10%. ···
- Discrete crystals are imaged directly at 0.1% crystalline naproxen.
- On the stored dispersion, all three methods show the same rising crystallinity and all show growing crystal signals. ···
- The selectivity rests on established SHG physics: isotropic material gives no net coherent signal.
- Linear scaling "in solid dispersions" rests on physical mixtures; for the dispersion there is only agreement over about 24–54%.
- Earlier detection in the dispersion is at about 24% of the drug (about 6 wt% of the powder), close to the PXRD limit, not at trace level.
- The 10 ppm LOD is a single-voxel calculation on a sample without naproxen; the lowest level imaged is 0.1%, and neither the intensities nor the background value are printed.
- "Actual crystallinity" is the weighed composition, not checked after cryomilling; all three methods were calibrated to the same nominal values.
- The stated spike recovery (about 93%) is 95.1% or 89.4% when recomputed, depending on the definition.
- SONICC lies between PXRD and Raman only from 4 h; at 2 h it is above both.
- The advantage of crystalline standards was not tested against dispersion standards, and blend and dispersion values use different crystallinity bases. ·
- Storage RH, container and the spiked sample's history are not printed.
- The text and tables disagree in places: 0–25% slope 0.92 vs 0.941; the abstract's 2.7% "prediction" error is the RMSEE; the Discussion cites Table 1 for Table 2 values; the PXRD current is printed as "30 mV". ···
- Laser-power linearity and particle-size response were not measured. ·
09Figure guide
Every figure and table of the paper, what it is there for, and what to check on the original. Key figures are shown with the story.
| Figure | What it does in the paper | Supports | Check (AI note) |
|---|---|---|---|
| Fig. 1 (p.14)key figure | Shows the instrument: an 800 nm, 100 fs, 80 MHz laser scanned by a resonant and a galvo mirror, with epi and transmission SHG detectors. | 0.2 | A schematic, not data. |
| Fig. 2 (p.15)key figure | The first sign of hidden crystals: the fresh dispersion gives a PXRD halo and no 863 cm⁻¹ Raman peak, yet SONICC shows sparse crystals; the 75/25 physical mixture is crystalline by all three. | F4, K4 | No independent check of the crystalline content; SONICC later reads about 24% of the drug, about 6 wt% of the powder. |
| Fig. 3 (p.16)key figure | The core calibration: SHG crystallinity against weighed content in the blends, 0–100% with a 0–25% insert, on a straight line with narrow 95% prediction bands. | F1, K1 | "Actual" crystallinity is the weighed composition, not checked after milling. |
| Fig. 4 (p.17)key figure | Sensitivity in pictures: 3D images of 100% crystalline naproxen and of 0.1% in 99.9% HPMCAS (about six discrete crystals), with the depth profile of one crystal. | F3, K3 | The 10 ppm LOD is calculated; 0.1% is the lowest level imaged. |
| Fig. 5 (p.18)key figure | The reference methods on the same blends: PXRD (a) and Raman (b) OPLS calibrations, with wider bands for PXRD and narrower ones for Raman than for SONICC.· | F2, K2 | Calibrated to the same nominal compositions as SONICC. |
| Fig. 6 (p.19)key figure | Crystallization during 60 °C storage seen three ways: a first PXRD peak near 19° 2θ at 2 h, a growing 863 cm⁻¹ Raman shoulder, and more SHG spots at 2 h than at 0 h. | F6, K4, K5 | Storage RH and container not printed. |
| Fig. 7 (p.20)key figure | The kinetics comparison: crystallinity against time by SONICC from 0 h and by PXRD and Raman from 2 h, all rising and incomplete at 48 h. | F6, K5, K7 | SONICC lies between the other two only from 4 h; no regression; SONICC used a blend standard, the others dispersion standards.·· |
| Table 1 (p.21) | Calibration statistics on the blends: slope, R² and RMSEE for SHG (0–100 and 0–25%), PXRD and Raman. | F1, F2, K1, K2 | The text gives slope 0.92 for 0–25%; the abstract's 2.7% "prediction" error is the RMSEE; the title omits the Raman row.·· |
| Table 2 (p.22) | The independent test: three prediction-set samples (3.5, 18 and 83%) read by SHG, PXRD and Raman. | F2, K2 | The Discussion cites these values as Table 1. |