Can FT-Raman spectroscopy quantify the degree of crystallinity, and so the amorphous content, of a model drug across the whole composition range, including small amounts of either form?
Yes, for physical mixtures of crystalline and amorphous indomethacin: a band-ratio calibration was linear from 0 to 100% crystalline, independent mixtures fell on a line of slope 1.008, about 1% of either form was detected, and the method RSD at 55% crystalline was 7.3%, set mainly by mixing inhomogeneity. ·····
For physical mixtures of γ-crystalline and quench-cooled amorphous indomethacin, measured by 1064 nm FT-Raman with a laser spot of about 1 mm, the ratio of the 1698 to the 1680 cm⁻¹ carbonyl band, corrected for the overlap of the two bands, gave a calibration that the authors describe as linear across the entire composition range and that independent mixtures followed (slope 1.008, intercept −0.0139; "reasonably rugged"). ··· They judge that about 1% amorphous or crystalline content can be detected ("would seem reasonable"), but that their estimated quantitation limit of about 2% is "too low". · Inhomogeneous mixing within the powder samples, rather than laser power, instrument fluctuation or separate sample preparation, is named the largest source of error, and laser heating did not crystallise the amorphous component "in this instance". ·· Milled crystalline indomethacin gave spectra like those of the mixtures, and its crystallinity fell with milling time, "estimated to be equivalent" to mixtures with 85.0, 77.5 and 63.5% crystalline after 5, 15 and 30 min; the effect of particle size was not evaluated, and correlation with other techniques for processed samples is left to further work. ··· The authors call FT-Raman "a potentially useful method" to complement existing techniques, on the basis of what they call a "preliminary study".
01Why this study
XRD, moisture sorption and microcalorimetry are the usual quantitative methods; methods for below 5% amorphous, and for in situ or on-line use, are sought.
Saleki-Gerhardt 1994, ref 2; refs 3–5
XRD detects around 10% amorphous material; water sorption and isothermal microcalorimetry about 1% disorder.
Saleki-Gerhardt 1994, ref 2; Briggner 1994, ref 16
It is fast and needs no sample preparation; amorphous content was "in theory" within reach.
Langkilde 1997, ref 6; refs 7, 8
The amorphous form is easily made by quench cooling the melt.
Taylor 1997, ref 9
A two-band model handles peaks that are not fully resolved.
Kontoyannis 1997, ref 13
More detail
What was already known.
- Why disorder matters. Milling, compaction and other processing can introduce disorder into crystalline materials, from local surface disorder to complete disruption of the lattice. Amorphous material may raise solubility and hygroscopicity, change mechanical and flow properties and lower chemical stability; yet the amorphous state is sometimes wanted, for the higher initial solubility of poorly soluble drugs or for sugar lyoprotectants in freeze-dried proteins. The introduction gives no reference for these points. ·
- Current methods. Many techniques detect and characterise the amorphous state (Hancock 1997, ref 1; Saleki-Gerhardt 1994, ref 2). For quantitation, XRD, moisture sorption and microcalorimetry are the most used (refs 2–5). Methods for low amorphous levels (below 5%) and for in situ and on-line testing are of particular interest. In the discussion the authors add that XRD measures the average degree of crystalline order with a detection limit of around 10% amorphous content, whereas water vapour sorption and isothermal microcalorimetry, which rely on the higher energy of the disordered phase, detect as little as 1% (Saleki-Gerhardt 1994, ref 2; Briggner 1994, ref 16); IR and XRD crystallinities correlate well (Black 1977, ref 4). ·
- FT-Raman for solid forms. FT-Raman has quantified mixtures of polymorphs using distinct peaks (Langkilde 1997, ref 6) or factor analysis (Deeley 1991, ref 7; Tudor 1993, ref 8). It is rapid and needs no sample preparation, and in theory it could quantify amorphous content if the amorphous and crystalline spectra differ enough, as ref 6 suggested.
- Indomethacin. The group had characterised crystalline and amorphous indomethacin by FT-Raman (Taylor 1997, ref 9). Indomethacin is poorly water soluble, is easily made amorphous by quench cooling the melt, and its amorphous and stable γ forms can be told apart by FT-Raman.
- Overlapping bands. Kontoyannis et al. quantified calcium oxalate hydrates by Raman although their characteristic peaks are not completely resolved, by including the contribution of each peak to the other (Kontoyannis 1997, ref 13). ·
- Error budget. The error parameters tested follow in part Bugay et al., who examined errors in polymorph quantitation by diffuse reflectance IR and XRPD (Bugay 1996, ref 10).
- Two-state and one-state disorder. Physical mixtures are a two-state model in which each particle is wholly amorphous or wholly crystalline, whereas mechanically processed material is likely to be partially crystalline throughout, a one-state model (Hüttenrauch 1985, ref 14).
What the paper set out to do. Prepare mixtures of amorphous and γ-crystalline indomethacin in known ratios, derive an equation that gives percentage crystallinity from measured Raman intensity ratios, and use it to estimate crystallinity in validation mixtures and in milled samples. ··
02How they approached it
One compound in two pure forms gives standards of exactly known crystallinity. One carbonyl band per form, corrected for their overlap, turns a band ratio into a crystalline fraction. Independent mixtures test the calibration, a 55% crystalline mixture carries the error budget, and milled samples test processed material. ··
What was done
Melt-quench the γ form→XRPD and DSC of both forms→100% crystalline and 100% amorphous references.·····
FT-Raman of the pure forms→carbonyl bands at 1698 and 1680 cm⁻¹→K ratios 0.052 and 0.401.··
Two-band intensity model with each band's contribution to the other→linear plot whose slope is the remaining K ratio.··
Calibration mixtures, 0–100% crystalline→band ratio I_R→slope 2.72→Eq. 5 for the crystalline fraction.···
Independent mixtures, 1–99% crystalline→Eq. 5→validation line and Table 1.··
55% crystalline mixtures→laser power, repeat scans, 6 samples × 5 shaken scans→error sources and RSD.···
SD of the pure forms→LOD and LOQ.·
Crystalline indomethacin milled 5, 15 and 30 min→Raman→crystallinity by Eq. 5.··
More detail
What was done
- Reference materials. γ-indomethacin from Sigma, a fine powder passing a 150 µm sieve, was assumed to be 100% crystalline. Amorphous indomethacin was made by melting it at 165 °C for 5 min, quench cooling in liquid nitrogen, warming to room temperature under vacuum, lightly pulverising to below 150 µm and storing over phosphorus pentoxide at −20 °C. XRPD (10–40° 2θ, 0.05° steps, 1.5 s per step) and DSC (pinhole pans, 20 °C/min, dry nitrogen) characterised both. ·····
- Mixtures. Known ratios were mixed geometrically to 500 mg in a Wig-L-Bug stainless steel capsule without the ball, transferred whole to a sealed glass vial of about 1 cm diameter and stored at −20 °C and 0% RH. Agitating the crystalline powder in the capsule did not change its DSC enthalpy of fusion. ··
- Raman measurement. A Bruker RFS 100 with a 1064 nm Nd:YAG laser, focused to a spot of about 1 mm (the largest the instrument allows) at about 450 mW, collected 180° backscattered Stokes spectra with a liquid-nitrogen-cooled germanium detector; 100 scans over 4000–50 cm⁻¹ at 4 cm⁻¹ resolution were averaged, with the powder in a sealed thin-walled glass vial.
- Quantitation. The peak heights at 1698 and 1680 cm⁻¹ gave the intensity ratio I_R. The pure-form spectra gave two K ratios (means of 10 measurements); the calibration mixtures gave the third from the slope of the Fig. 4 plot; Eq. 5 then converts any I_R into a crystalline fraction. ···
- Validation. Twelve mixtures not used for the calibration, from 1 to 99% crystalline, were measured in 5 runs per sample, with 1, 3 or 6 independent samples per composition. ·
- Errors. At 55% crystalline: three consecutive scans at each of 225, 450 and 900 mW; 10 consecutive scans without moving the sample; and 6 independent samples, each scanned 5 times with removal and shaking in between, for the homogeneity range and the overall RSD (Eq. 1). LOD and LOQ were taken as 3 and 10 SD of the crystallinity measured 10 times on a pure form. ······
- Milling. 500 mg of crystalline indomethacin was milled in the same capsule with a metal ball for 5, 15 and 30 min and measured like the standards (5 determinations each). ·
Key equations
Each measured band height is the sum of the crystalline and amorphous contributions at that position, weighted by the fraction x of each form; K is a band factor that includes the instrument response and the scattering strength. This is how the paper handles the broad amorphous band adding to the crystalline one.·
The two ratios measured on the pure forms (0.052 and 0.401) are inserted; plotting (x_c − 1)(I_R − 0.401) against x_c(0.052 I_R − 1) for the calibration mixtures then gives the last ratio, K_c^1698/K_a^1680 = 2.72, as the slope.···
Inserting the slope of 2.72 into Eq. 4 gives this calibration equation, which turns one measured band ratio into a crystalline fraction.·
Used for the overall method error at 55% crystalline (7.3%).·
Terms used on this page
- Degree of crystallinity: the fraction of the sample that is crystalline, written x_c (the mole fraction of the crystalline form); 1 − x_c is the amorphous content.
- Two-state and one-state model: in a physical mixture every particle is either fully crystalline or fully amorphous (two-state); after milling, particles may be partly disordered throughout (one-state).
- Limit of detection and limit of quantitation: the smallest content that can be told apart from zero (here 3 SD of the reading on a pure form) and the smallest that can be measured with acceptable precision (here 10 SD).
- RSD: the standard deviation as a percentage of the mean.
- Benzoyl carbonyl band: the C=O stretching vibration of indomethacin's chlorobenzoyl group; its position differs between the crystal and the glass.
03Main findings
Each form has its own carbonyl band, and the two overlap
Crystalline indomethacin has a narrow carbonyl band at 1698 cm⁻¹, amorphous a broader one at 1680 cm⁻¹; in mixtures they are only partly resolved. The crystalline band adds little at 1680 (ratio 0.052), the amorphous band more at 1698 (ratio 0.401). ···
More detail
XRPD showed sharp peaks for the γ form and only a broad halo for the quench-cooled glass; DSC gave a Tg of 45 °C for the glass and, for the crystal, no Tg and fusion at 161 °C with 109 J/g. The authors call this "excellent agreement" with earlier work and take the two materials as 100% amorphous and 100% crystalline. ··· The full FT-Raman spectra of the two forms differ over the whole range (Fig. 2), and the authors chose the benzoyl carbonyl bands: 1698 cm⁻¹ for the crystal and 1680 cm⁻¹ for the glass. ·· In mixtures with 20, 50 and 80% amorphous content the 1680 cm⁻¹ band grows from a weak tail to a shoulder to a maximum of about the same height as the 1698 cm⁻¹ band (Fig. 3). The bands are not completely resolved, which the authors put down to the broadening of the amorphous carbonyl band, so the contribution of each band to the other was included by the two-band method of Kontoyannis et al. · From the pure-form spectra (10 measurements each), the crystalline band has 0.052 of its 1698 cm⁻¹ height at 1680 cm⁻¹, while the amorphous band has 0.401 of its 1680 cm⁻¹ height at 1698 cm⁻¹. The authors read the small value as a narrow crystalline peak with little overlap and the larger one as a broad amorphous peak that adds more to the crystalline peak in mixtures. ·
A band-ratio calibration is linear from 0 to 100% crystalline
Calibration mixtures fell on a straight line in the linearised form (slope 2.72, r = 0.993), which fixed Eq. 5. Recalculated with Eq. 5, the same mixtures gave slope 1.01, intercept −0.01, r = 0.9996 against their known crystallinity. ··
More detail
I_R was measured on mixtures covering 0–100% crystalline. Plotted as (x_c − 1)(I_R − 0.401) against x_c(0.052 I_R − 1), twelve points lay close to a least-squares line with correlation coefficient 0.993 and slope 2.72, which is the ratio K_c^1698/K_a^1680 (Fig. 4). · Inserting it into Eq. 4 gave Eq. 5, x_c = (I_R − 0.401)/(0.859 I_R + 2.316). · Recalculating the calibration mixtures with Eq. 5 gave values within about 0.03 of their known fractions from about 0.05 to 1.0 (read from Fig. 5) and a line of slope 1.01, intercept −0.01 and correlation coefficient 0.9996, which the authors present as "demonstrating the validity of equation 5". ·
Independent mixtures follow the calibration, down to about 1% of either form
Twelve mixtures not used for the calibration fell on a line of slope 1.008 and intercept −0.0139; 99% crystalline read 98.7% and 1% read 0.9%. Between 2 and 25% crystalline the means read 0.8–3.7 points low. ···
More detail
The validation mixtures spanned 1 to 99% crystalline, each measured in 5 runs on 1, 3 or 6 independent samples (Table 1). The measured means were 98.7, 97.7, 96.5, 93.5 and 96.3% for 99, 98, 97, 93 and 96% crystalline; 73.6% for 75; 54.0% for 55; 21.3% for 25; 5.9% for 7; 2.2% for 3; 1.0% for 2; and 0.9% for 1, with SDs from 0.2 to 4.0 points. Plotted against the actual values they gave a line of slope 1.008 and intercept −0.0139 (Fig. 6), which the authors read as a "reasonably rugged" assay with calculated and input values "in good agreement". · They conclude that small amounts of both forms, as well as mid-range compositions, can be detected and quantified ("It is apparent"). Because the ranges are wide, they say several measurements "must be averaged", and that accuracy improves from 5 measurements (one sample) to 15 (three samples) or 30 (six samples). At 25, 7, 3 and 2% crystalline the whole 5-run range lies below the actual value; from 93 to 99% crystalline the means are within 0.5 points of it. ·
Mixing, not the instrument, sets the error
At 55% crystalline, laser powers of 225–900 mW gave 54.4–56.6% and 10 repeat scans 55.8–56.3%, but 30 scans of 6 shaken samples ranged from 44.8 to 62.5% (mean 54%, SD 4%, RSD 7.3%). ···
More detail
Raising the laser power from 225 to 900 mW changed the band ratio very little, although the absolute intensities rose: the crystallinity stayed within 54.4–56.6%. · Ten consecutive scans of one sample spanned only 55.8–56.3%. The authors read the stability under higher power and repeated scanning as showing that laser heating did not crystallise the amorphous component "in this instance". In contrast, 6 independently prepared 55% mixtures, each scanned 5 times with shaking between scans, gave values from 44.8 to 62.5% (Table 2). The spread among the 5 scans of one sample (SDs 1.4–5.3 points) was comparable to that among the 6 samples at a given scan (SDs 2.7–4.5), and the sample means (51.9–56.1%) were close, so the authors attribute the scatter to non-homogeneous mixing within a sample rather than to differences between preparations. · All 30 values gave a mean of 54%, an SD of 4% and an overall method RSD of 7.3% at this composition; because the SDs of Table 1 show no trend with crystallinity, the authors expect the RSD to depend on composition. · Errors in measuring the peak heights contribute to all of these.
Milling lowers the Raman-estimated crystallinity
After milling for 5, 15 and 30 min, crystalline indomethacin gave spectra like those of the mixtures, with more intensity at 1680 cm⁻¹, and read as 85.0, 77.5 and 63.5% crystalline. ··
More detail
Milling raised the intensity at 1680 cm⁻¹, which the authors take as the formation of amorphous material; the spectra otherwise resembled those of the calibration mixtures, with no change in peak positions. · By Eq. 5 the crystallinity fell with milling time and was "estimated to be equivalent" to binary mixtures with 85.0% (range 85.3–84.8), 77.5% (78.5–76.0) and 63.5% (61–65.5) crystalline after 5, 15 and 30 min (means of 5 determinations). · The authors note that milled material is likely to be partially crystalline throughout (one-state) rather than a mixture of fully crystalline and fully amorphous particles, and that milling also reduces particle size, which can affect Raman intensity and was not evaluated. ·
04Further observations
- Detection limit 0.6% and quantitation limit about 2%, from a pure-form SD of 0.2%; the authors find 1% detection reasonable but think the 2% quantitation estimate is too low. ·
- Mixing in the capsule did not change the crystal's enthalpy of fusion. ·
More detail
- Detection and quantitation limits. Ten measurements of pure crystalline indomethacin gave a crystallinity SD of 0.2%, so LOD = 3 SD = 0.6% and LOQ = 10 SD, about 2%; pure amorphous indomethacin also gave an SD of 0.2%, so similar limits apply to crystalline content. Table 1 makes a detection limit of around 1% "seem reasonable", fairly close to the estimate, but the variation of the mixture data "suggests" that the quantitation estimate is too low. ···
- Mixing does not alter the crystal. Agitating crystalline indomethacin in the Wig-L-Bug capsule without the ball did not change its DSC enthalpy of fusion (values not printed). ·
05How the authors connect the findings
More detail
- The γ crystal and the glass each show a benzoyl carbonyl band at a distinct position, 1698 and 1680 cm⁻¹, and these were chosen for quantitation. ·
- The amorphous band is broadened, so the two bands are not completely resolved; the crystalline band has only 0.052 of its height at 1680 cm⁻¹, but the amorphous band 0.401 of its height at 1698 cm⁻¹. The model must therefore count both contributions. ··
- With the Kontoyannis two-band model (Eq. 3), the two pure-form ratios and the slope of 2.72 from the calibration mixtures, Eq. 5 gives the crystalline fraction from one measured ratio. The K ratios are taken to depend only on the scattering parameter of each band. ···
- Independent mixtures fall on a line of slope 1.008 and intercept −0.0139, read as a "reasonably rugged" assay; about 1% of either form can be detected. ···
- Laser power and repeated scanning change the result by at most 2.2 points, while re-sampling the same powder spreads it over 17.7 points (44.8 to 62.5%); the authors attribute the scatter to mixing inhomogeneity within a sample. ···
- Raman samples a small area of the powder bed, and the representativeness of a scan depends on particle size and homogeneity: large particles close to the spot size, wide size ranges or clusters of one component give non-representative scans. In milled samples a uniform disorder per particle would not matter, but a distribution of disorder that depends on particle-size reduction is "more likely".
What sets the precision. The sampled area relative to the scale of inhomogeneity. Re-mixing by shaking between scans was not efficient compared with a rotating sample holder (ref 6), and an averaging device "would improve this method substantially" and lower the detection and quantitation limits. ···
Conclusion drawn. "This preliminary study suggests" that Raman spectroscopy is a useful method for quantifying crystallinity that can complement XRD, sorption and calorimetry; for processed samples, correlations with existing techniques still have to be established. ·
06What it means for formulation
- FT-Raman can complement XRD and sorption or calorimetry for crystallinity, with about 1% detection here. ··
- Calibration standards should match the particle size of the test samples.
- On-line monitoring of milling by FT-Raman is conceivable.
- For a drug whose crystal and glass have distinct bands, a two-band ratio calibration from physical mixtures can reach about 1% detection without sample preparation.
- Sampling, not spectroscopy, limits precision: average over many spots before trusting low amorphous levels.
More detail
- Raman, like IR, gives molecular information on both phases without transforming them and needs no sample preparation; the authors see it as a complement to XRD, which detects around 10% amorphous material, and to water sorption and isothermal microcalorimetry, which detect around 1% disorder. ···
- Laser heating could crystallise an amorphous sample if care is not taken; cooling accessories may be used.
- To reduce sub-sampling, spectra should be averaged over several regions, for example with a rotating sample holder; an acetone slurry improves mixing but needs caution with amorphous materials that may sorb solvents. ·
- Calibration standards should match the particle-size distribution of the test samples, or its influence should be evaluated, and the particle sizes of both powders in binary standards should be controlled.
- FT-Raman could conceivably monitor pharmaceutical processes on line, for example the effect of a milling step on polymorphic form or crystallinity, helped by the weak Raman scattering of polymeric excipients and by fibre-optic probes. ·
- A two-band calibration built from a crystalline and a quench-cooled amorphous reference is a simple template for quantifying low amorphous content in a drug substance, provided the two forms have distinct bands; this paper sets an early benchmark of about 1% detection for indomethacin.
- The 7.3% RSD at 55% crystalline and the low readings at 2–25% crystalline mean that single-spot readings should be averaged and reported with their spread.
- Raman-derived crystallinity of milled material is an equivalent on a physical-mixture scale; it may differ from XRPD or calorimetry when disorder is spread through the particles.
07Open questions
- How does particle size affect the Raman intensity of these powders? ·
- Does Raman crystallinity of processed samples agree with other techniques?
- The milled samples were not checked by XRPD or DSC.
- Quantitation below about 25% crystalline reads low, and the 2% row of Table 1 is internally inconsistent. ·
More detail
- The effect of particle size, which milling reduces, on Raman intensity was not evaluated; for inorganic materials, intensity rises as particles get smaller, with changes of the order of 100 µm giving observable differences. ·
- For processed samples, correlations with existing techniques, which may respond to different properties of the amorphous or crystalline state, remain to be established.
- A sampling device that averages over the sample is expected to improve precision and the detection limits; it was not tested.
- The milled samples were not measured by XRPD, DSC or calorimetry, so their Raman-equivalent crystallinities are not tested against another method.
- Between 2 and 25% crystalline the readings run low, and the 2% row of Table 1 cannot be internally consistent, so quantitation at the low end is less certain than the abstract's 1% detection suggests. ·
- Laser heating was tested only on a 55% crystalline mixture, without measuring the sample temperature.
- The gain from averaging 15 or 30 measurements is shown only at 55% crystalline.
08How solid
AI assessment, not a statement made in the paperThe calibration and its independent validation on physical mixtures are direct and consistent, and the error sources are measured separately. But the low end reads systematically low, several table values are inconsistent, and the milled-sample values rest on the mixture calibration alone. ·····
More detail
- The two reference forms are characterised by XRPD and DSC. ··
- The validation uses mixtures not used in the calibration, across 1–99% crystalline, and agrees with it (slope 1.008). ·
- The error sources are separated experimentally: laser power and repeat scanning spread the result by 2.2 and 0.5 points, re-sampling the powder by 17.7 points (44.8 to 62.5%) at 55% crystalline. ··
- The calibration equation is reproduced from the printed ratios within rounding.
- Fig. 5 replots the calibration mixtures that set the slope, so it shows internal consistency rather than validity.
- At 2–25% crystalline the means are 0.8–3.7 points low, with ranges that exclude the true value; from 93 to 99% they are within 0.5.
- Some table values are inconsistent: the 2% row's mean, SD and range cannot all hold; a Table 2 mean is misprinted (3.1 for 53.1); the RSD recomputes to 7.2%. ··
- The milled-sample values are equivalents on a two-state calibration, with particle size not evaluated and no second technique.
- The benefit of averaging and the absence of laser heating effects rest on 55% crystalline data only. ·
- The stated validation slope and intercept are reproduced only approximately from Table 1.
09Figure guide
Every figure and table of the paper, what it is there for, and what to check on the original. Key figures are shown with the story.
| Figure | What it does in the paper | Supports | Check (AI note) |
|---|---|---|---|
| Fig. 1 (p.756)key figure | Confirms the references: sharp peaks for the γ crystal, only a halo for the glass. | F1 | |
| Fig. 2 (p.757)key figure | Shows that the full Raman spectra of the two forms differ. | F1 | |
| Fig. 3 (p.757)key figure | Picks the carbonyl bands and shows their overlap in 20–80% amorphous mixtures. | F1 | |
| Fig. 4 (p.758)key figure | Linearised calibration; its slope (2.72) completes Eq. 5. | F2 | |
| Fig. 5 (p.758)key figure | Calibration mixtures recalculated with Eq. 5: slope 1.01, r = 0.9996. | F2, K1 | Same mixtures that set the slope; not an independent test. |
| Fig. 6 (p.758)key figure | Independent validation: slope 1.008, intercept −0.0139. | F3, K1 | Fit point set not stated; Table 1 means give 1.010 and −0.0134. |
| Table 1 (p.759)key figure | Measured against actual crystallinity with SD and range for 12 compositions. | F3, K2 | 2% row internally inconsistent; low end reads low.· |
| Table 2 (p.760)key figure | Inter- and intra-sample variation at 55% crystalline, the basis of the error budget. | F4, K3 | Scan-2 mean misprinted (3.1 for 53.1). |