Can the crystalline fraction of a drug in a partly crystallized amorphous solid dispersion be measured more accurately from PXRD by PLS regression than by the ratio of Bragg-peak area to total area?
Yes, for felodipine–PVP calibration samples: PLS gave RMSEE of 2.2, 1.9 and 4.7% drug crystallinity at 25, 50 and 75% PVP, against 11.2, 17.0 and 23.6% for the area method. These are errors on the calibration samples themselves; no independent test set was measured. ·
For felodipine–PVP calibration samples at 25, 50 and 75% PVP, PLS regression of the normalized PXRD pattern estimated drug crystallinity "far better" than the Bragg-peak-to-total-area ratio, with RMSEE of 2.2, 1.9 and 4.7% against 11.2, 17.0 and 23.6% drug crystallinity. · The area method grew worse as polymer content rose and the Bragg-peak share of the pattern fell. · At 75% PVP the authors read the 4.7% error as about 1.2% crystalline drug in the total mass, comparable with what a laboratory diffractometer detects in simple crystalline–amorphous blends ("indicates"). · They interpret the second PLS component as shifts of the Bragg peaks and the halo plus slight peak broadening, and "It is thought" that these subtle variations, from small crystallites or crystal strain and intimate drug–polymer mixing, explain the gain. · Two links were not measured: prediction of samples outside the calibration set, and the peak positions, widths or crystallite sizes behind the second component. ·
01Why this study
The drug in an amorphous dispersion sits above its solubility in the polymer, so it can crystallize during storage.
Detection limits of 5% crystallinity or lower were reported for amorphous–crystalline drug mixtures.
Nunes 2005, ref 21; Surana 2000, ref 30
The amorphous pattern of a molecularly mixed dispersion is not a sum of the pure amorphous patterns.
Rumondor 2009a, ref 24; Newman 2008, ref 20
More detail
The paper cites by author and year; ref N is the entry's position in the printed reference list (35 entries, pp.159–160).
What was already known.
- Amorphous dispersions as a solubility tool. Delivering a poorly soluble drug in the amorphous form, intimately mixed with a polymer, is "one potential method" to raise bioavailability. (Chiou and Riegelman 1971, ref 5; Leuner and Dressman 2000, ref 12; Serajuddin 1999, ref 26; Ford 1986, ref 9)
- The advantage lasts only while the drug stays amorphous. The free energy of the drug in a dispersion is typically higher than that of the crystal, so the drug is above its "solubility limit" in the polymer, crystallization cannot be retarded indefinitely, and it may proceed to varying degrees during storage; quantifying amorphous and crystalline drug is therefore desirable. (Tao 2009, ref 32; Marsac 2006b, ref 17; Marsac 2009, ref 15; Rumondor 2009b, ref 25)
- Crystallinity had been quantified in simpler systems. DSC, infrared and Raman spectroscopy and PXRD had been used for pure drugs and polymorph mixtures (for example Clas 1995, ref 6; Taylor and Zografi 1998, ref 33; Shah 2006, ref 28), and crystalline drug had been quantified in multicomponent mixtures (Suryanarayanan and Herman 1991, ref 31; Moore 2008, 2009, refs 18, 19). Quantification inside an otherwise amorphous, molecularly mixed dispersion had not been widely reported.
- PXRD is sensitive enough. PXRD is well suited to identifying crystalline material, with detection limits of 5% or lower for amorphous–crystalline drug mixtures. (Fix and Steffens 2004, ref 8; Nunes 2005, ref 21; Chen 2001, ref 3; Yoshioka 1994, 1995, refs 34, 35; Otsuka and Kaneniwa 1988, ref 22; Surana and Suryanarayanan 2000, ref 30) ·
- The area method. Crystallinity can be estimated as Bragg-peak area over total area if crystalline and amorphous intensities scale with the phase fractions. (Shah 2006, ref 28; Nunes 2005, ref 21) ·
- Felodipine–PVP is an intimately mixed model system. It is "known to" form a molecular-level amorphous dispersion. (Marsac 2006a, 2006b, 2008, 2010, refs 14, 17, 13, 16; Konno and Taylor 2006, ref 10)
- Mixing changes the halo. The X-ray pattern of a molecular-level dispersion cannot be described as a linear combination of the pure amorphous patterns; intimate mixing may change the halo shape. (Rumondor 2009a, ref 24; Newman 2008, ref 20)
What the paper set out to do. Develop a way to quantify drug crystallinity inside a molecular-level dispersion with PXRD, using calibration samples that contain the real dispersion, and test whether PLS multivariate modelling of the pattern beats the traditional peak-area method. ··
02How they approached it
One drug–polymer pair at three polymer levels, each with six known crystallinity levels in triplicate. The calibration samples were made from the real dispersion plus crystalline drug, and the same diffractograms were analysed by two methods. ···
What was done
FEL and PVP dissolved in ethanol–dichloromethane→rotary evaporation→X-ray amorphous dispersion→crystalline FEL added→cryo-milled 4 min→18 samples per PVP level.··
Same mix milled 2, 4 or 6 min→IR and PXRD→milling effect.···
PXRD 5–35° 2θ at 8°/min→normalized to the Si standard and to unit area.··
Bragg-peak area by Fourier-transform filtering over total area (Eq. 1)→calculated versus actual crystallinity.··
Two-component PLS of the whole pattern, log-transformed→R2Y, loading weights, calculated versus actual crystallinity.···
Area method against PLS at each PVP level: slope, R², RMSEE and 95% prediction bands.··
More detail
What was done
- Dispersions. Crystalline felodipine and PVP K29-32 (dried over P2O5 for at least a week) were dissolved at the desired ratio in 1:1 (w/w) ethanol:dichloromethane, rotary-evaporated, dried overnight under vacuum and gently ground below 10% RH; PXRD confirmed them X-ray amorphous. ···
- Calibration samples. Crystalline felodipine was added to the dispersion powder so that the overall PVP content was 25, 50 or 75% w/w and 0, 10, 25, 50, 75 or 90% of the drug was crystalline. Each mix was cryo-milled (two impacts per second, 4 min) and sieved through 150 µm. Three independent replicates were made per point, 18 samples per PVP level. Because the overall drug:polymer ratio is fixed, the amorphous part loses drug as the crystalline share rises. ·
- Milling control. Samples with 25% PVP and half the drug crystalline were milled 2, 4 or 6 min and checked by ATR-IR and PXRD.
- PXRD. Shimadzu XRD-6000, Cu-Kα, Bragg–Brentano geometry, 5–35° 2θ, 8°/min, 0.04° steps; the Si [1 1 1] peak was checked daily (28.423–28.463°), and three consecutive readings per sample were overlaid to check that no crystallization occurred during the scan.
- Pretreatment. Each pattern was normalized to the same-day Si-standard intensity and then to unit area.
- Area method. Total area by the trapezoidal rule; Bragg-peak area after Fourier-transform filtering; crystallinity by Eq. 1; actual versus predicted fitted by a straight line with 95% prediction bands. Variants (total Bragg-peak area, other normalizations) were also tried. ··
- PLS. SIMCA-P+ v.11, one model per PVP level, two components; normal probability plots showed nonlinearity, so log-transformed models were used (which variable was transformed is not stated). ···
Key equations
The Bragg-peak area A_c over the total area under the pattern (Bragg peaks plus amorphous halo A_a, 5–35° 2θ) gives the crystalline share, provided both intensities scale with the phase fractions.·
Crystallinity is defined per drug; multiplying by the drug mass fraction gives crystalline drug per sample mass, so the 4.7% error at 75% PVP is about 1.2% of the sample.·
Terms used on this page
- Bragg peaks: sharp diffraction peaks from crystalline material; amorphous halo: the broad hump from non-crystalline material.
- PLS (partial least-squares) regression: a multivariate method that compresses the whole pattern into a few latent components chosen to predict a property, here % crystallinity.
- Loading weights: how strongly each 2θ position contributes to a PLS component.
- R2Y: the fraction of the variation in crystallinity explained by a component (cumulative over components).
- RMSEE (root-mean-squared error of estimation): the typical error on the calibration samples, in % of drug crystallinity.
- Prediction band: the range within which 95% of new single predictions are expected to fall.
- Vainshtein's law: the total diffracted intensity of a sample is conserved whether its material is crystalline or amorphous.
03Main findings
The area ratio misjudges crystallinity, and more so as polymer rises
Calculated values ran from about −12 to 126% for actual levels of 0–90%. From 25 to 75% PVP, RMSEE rose from 11.2 to 17.0 to 23.6% and R² fell from 0.93 to 0.80 to 0.75. ·
See also Table 1
More detail
In the normalized patterns, Bragg peaks of crystalline felodipine (strongest near 10.3, 16.5 and 23.3° 2θ) grow out of the broad halo of the dispersion as the crystalline share rises from 0 to 90%; at 50% PVP the 10.3° peak is just visible at 10% crystallinity (Fig. 1). · Eq. 1 turned each pattern into one number. Against the known crystallinity the results scattered widely (Fig. 2, read from the plot): at 25% PVP the 90% samples came out at about 108–126% and one 10% sample at about 47%; at 75% PVP the 0% samples ranged from about −12 to 14% and the 75% samples from about 47 to 119%. The 95% prediction bands lay about ±21, ±39 and ±46 points from the fitted line at 25, 50 and 75% PVP. Table 1 gives R² of 0.93, 0.80 and 0.75, RMSEE of 11.2, 17.0 and 23.6% and slopes of 1.26, 1.03 and 1.25. The authors read the falling R² and rising RMSEE as the result of a shrinking Bragg-peak share at higher polymer content, and the slopes away from 1 as "poor overall correlation". · Other area-based calculations, such as total Bragg-peak area or other normalizations, did not help: linearity was lost and RMSEE exceeded 20% at both ends of the crystallinity range (data not shown). · How RMSEE was computed is not stated; the raw scatter of the Eq. 1 values read from Fig. 2a is about 26%, so the tabulated 11.2% appears to be scatter about the fitted line. The Table 1 slopes are predicted-on-actual, the reciprocals of the lines drawn in Fig. 2.
PLS cuts the error five- to nine-fold
Two-component PLS models gave RMSEE of 2.2, 1.9 and 4.7%, R² of 0.996, 0.997 and 0.978 and slopes of 0.998 to 0.984, with much narrower prediction bands. ·
More detail
For each PVP level a PLS model related the whole normalized pattern to the known crystallinity; two components fitted the data "without over-fitting", and a log transformation removed nonlinearity seen in normal probability plots, mostly at 25 and 50% PVP. ·· Calculated values now lay close to the actual levels (Fig. 6, read from the plot): at 25% PVP the 90% samples came out at about 91–92% and the 0% samples at about −3 to 1%; the widest spread was at 75% PVP, where the 75% samples came out at about 63–76%. The prediction bands lay about ±6, ±6 and ±10 points from the line, against about ±21 to ±46 for the area method; the 75% PVP band was the widest, "most likely" because its Bragg peaks are weaker. ·· RMSEE fell from 11.2 to 2.2% at 25% PVP, from 17.0 to 1.9% at 50% and from 23.6 to 4.7% at 75%, which the authors call "far better" (abstract) and "much better" (results), with the gain "particularly apparent" at 75% PVP. ·· At 75% PVP the drug is 25% of the mass, so 4.7% of the drug is about 1.2% of the sample; the authors say this "indicates" that PLS can estimate crystallinity at that level, comparable with detection in blends of pure crystalline and amorphous material on a laboratory diffractometer (Surana and Suryanarayanan 2000, ref 30). ·· These RMSEE values are errors on the 18 calibration samples of each model; no cross-validation statistic or independent test set is reported.
The first PLS component weighs Bragg peaks against the halo; the second picks up finer changes
The first component explained R2Y of 0.872, 0.948 and 0.511 of the crystallinity variation at 25, 50 and 75% PVP; the second added 0.119, 0.0465 and 0.448.
More detail
The first-component loading weights follow the crystalline felodipine pattern, with maxima at its Bragg positions (Figs. 3–5). The authors report the halo contribution as negative and the Bragg contribution as positive: as crystals replace amorphous drug at a fixed drug:polymer ratio, peak intensity rises while halo intensity falls, so the total diffracted intensity is conserved, "in agreement with" Vainshtein's law. ··· The second-component loadings are noisy traces with sharp features at the main Bragg positions (near 10.3, 16.5 and 23.3° 2θ). The authors interpret them as shifts of the Bragg peaks and the halo and slight broadening of the Bragg peaks, and offer three "potential explanations": experimental inaccuracies, lattice strain and small crystallites possibly introduced by cryo-milling, and intimate drug–polymer mixing in the dispersion. ··· The second component carries most weight at 75% PVP (R2Y 0.448) and least at 50% PVP (0.0465). The loadings are printed normalized and offset, without a zero line, so their signs and the shift shapes cannot be checked on the figures.
04Further observations
- Milling for 2 or 6 min changed calculated crystallinity by less than 2%.
- Milling created no new dispersion by IR but lowered the Bragg peaks slightly. ·
- Slow scans (2 or 4°/min) made Bragg peaks grow between consecutive readings of nifedipine mixtures.
- No other area-based calculation beat Eq. 1.
- At 75% PVP the two PLS components share the fit almost equally.
More detail
- Milling does not distort the calibration. IR spectra of samples milled 2, 4 and 6 min overlapped, the felodipine–PVP interaction peaks did not grow and the crystalline felodipine band did not change; the authors take this to "indicate" that milling did not form detectable new dispersion. The diffractograms were almost identical, apart from a slight drop in Bragg-peak height that may reflect crystal defects. The calculated crystallinity of samples milled 2 and 6 min differed by less than 2%, "well within" the quantification errors. ·····
- Slower scans were not an option. Scanning mixtures of amorphous and crystalline nifedipine at 2 or 4°/min made the Bragg peaks grow from one reading to the next (data not shown), which the authors read as X-ray-induced crystallization ("suggest"); so the signal-to-noise ratio could not be raised by scanning longer. ··
- Other area-based calculations failed too. Correlating crystallinity with total Bragg-peak area, or using other area normalizations, did not improve the prediction; linearity was lost and RMSEE exceeded 20% at both extremes (data not shown). ·
- Nonlinearity needed a transformation. The first PLS models showed nonlinearity in normal probability plots, mostly at 25 and 50% PVP; log-transformed models were better.
- The weight of the second component differs by polymer level. It adds R2Y 0.119 at 25% PVP, 0.0465 at 50% and 0.448 at 75%, where the first component alone explains only 0.511.
05How the authors connect the findings
More detail
- The area method rests on the assumption that measured crystalline and amorphous intensities are proportional to the crystalline and amorphous fractions; areas rather than peak heights were used to reduce the effect of lattice strain and particle size. ·
- In a molecular-level dispersion the halo is not a sum of the pure amorphous halos, so its shape changes with mixing; this is why the calibration samples contain real dispersion plus crystals. The authors add that cryo-milling may introduce strain and small crystallites, which shift and broaden peaks. ···
Area ratio. Each pattern is reduced to one area ratio → the halo and peaks must be separated by Fourier-transform filtering → poor signal-to-noise ("One possible cause"), non-uniform samples and different proportionality constants for the two phases ("may have introduced errors") add scatter → RMSEE 11.2–23.6%, rising as the Bragg-peak share of the pattern shrinks with more polymer. ···
PLS. The whole pattern enters the model → the first component captures the main exchange of intensity between Bragg peaks and halo, as expected from Vainshtein's law → the second component captures shifts of the Bragg peaks and halo and slight peak broadening → RMSEE 1.9–4.7%. ···
What limits accuracy. The share of the pattern that comes from Bragg peaks. The area error rises as polymer content rises, and even with PLS the 75% PVP prediction band is the widest, "most likely as a result of lower sensitivity" from weaker Bragg peaks. ·
Conclusion drawn. PLS predictions were "much better" than those of Eq. 1, and "It is thought" that the improvement comes from subtle variations in the patterns, from small crystallites or crystal strain and intimate drug–polymer mixing, captured in the second-component loading weights. ·
06What it means for formulation
- Calibration samples for a dispersion should be the dispersion itself plus crystalline drug, not amorphous drug, crystals and polymer mixed separately.
- PLS needs less subjective and less time-intensive analysis, since halo and peaks need not be separated.
- For stability studies of a dispersion, a PLS calibration built at the same overall drug loading from the real dispersion plus crystals can track a few percent of crystalline drug, but it must be validated on samples it was not fitted to.
More detail
- The calibration samples must capture the multiphase nature of a partly crystallized dispersion: dispersion plus crystalline drug, because mixing changes the halo. ·
- PLS requires less subjective and time-intensive data analysis, since it works on the whole pattern without separating halo and Bragg-peak contributions.
- A PLS model also lets the analyst spot outliers and nonlinearity and see from its loadings which pattern features matter.
- Scanning slower to gain signal can itself crystallize an amorphous sample ("suggest").
- The reported 1.9–4.7% are calibration-fit errors, not detection limits; a cross-validated error or an independent test set would be needed before using the method to release or reject a batch.
- Crystallinity here is per drug. At low drug loading the same per-drug error is a much smaller mass of crystals, which matters when comparing with methods that report per sample.
- The calibration mimics crystallization from a dispersion, where the remaining amorphous phase loses drug; samples whose amorphous phase changes in a different way (for example by water uptake or phase separation) lie outside the calibration.
- The calibration crystals were milled; crystals grown slowly during storage may be larger and less strained, which the second component may treat differently.
07Open questions
- Which source causes the second-component variation: experimental shifts, strain and small crystallites from milling, or intimate mixing? ··
- How can signal-to-noise be raised without X-ray-induced crystallization? ·
- No independent test set, cross-validation or detection limit.
- One drug–polymer pair, milled calibration crystals only.
- How much of the gain needs the second component.
More detail
- The shifts and broadening in the second component have "several potential explanations": experimental inaccuracies, lattice strain and small crystallites possibly introduced by cryo-milling, and intimate mixing; none was tested. ··
- Poor signal-to-noise is "one possible cause" of the area-method errors, but slower scans risk X-ray-induced crystallization, so the trade-off is left open. ·
- How the PLS models perform on samples they were not fitted to; no cross-validation, test set, limit of detection or limit of quantitation is reported.
- Whether a one-component PLS model would already beat the area method; at 50% PVP the first component alone explains R2Y 0.948, above the area-method R² of 0.80.
- Whether the models work on crystals grown during storage, which were not milled.
- Only felodipine–PVP, one PVP grade, one instrument and one milling protocol were studied; transfer to other drugs, polymers or diffractometers was not tested.
08How solid
AI assessment, not a statement made in the paperThe calibration gain is large, consistent at all three polymer levels and directly tabulated. But the errors are fit errors on the calibration set, key modelling choices are unstated, and the explanation for the gain is untested. ···
More detail
- The comparison is direct: the same diffractograms analysed two ways, with slope, R² and RMSEE tabulated for both at three polymer levels. ·
- Six crystallinity levels with three independent replicates each give 18 samples per model.
- The plotted PLS predictions lie close to the actual levels at every PVP content.
- The milling control shows that the preparation step barely changes calculated crystallinity. ··
- Patterns were normalized to a daily Si standard and to unit area. ·
- RMSEE is an error of estimation on the calibration samples; no validation was done, and the 1.2% figure is not a detection limit. ·
- How RMSEE was computed for the area method is not stated; the raw Eq. 1 errors appear larger than tabulated.
- The log transformation, the component-selection criterion and the difference between Table 2 R2Y and Table 1 R² are not explained. ·
- The second-component explanation is not tested; its weight differs widely by PVP level, and the loadings cannot be checked for sign or shift shape. ·
- The amorphous part's composition changes with crystallinity inside each model, and the calibration crystals were milled. ·
- Outlier detection is claimed but not shown.
09Figure guide
Every figure and table of the paper, what it is there for, and what to check on the original. Key figures are shown with the story.
| Figure | What it does in the paper | Supports | Check (AI note) |
|---|---|---|---|
| Fig. 1 (p.157)key figure | Shows the calibration principle: Bragg peaks rise against the halo as crystalline felodipine is added at 50% PVP. | 0.2 | |
| Fig. 2 (p.157)key figure | Area-method predictions against actual crystallinity with 95% prediction bands at three PVP levels.· | F1, K1 | Table 1 slopes are the reciprocals of the drawn lines; raw errors look larger than the RMSEE.· |
| Table 1 (p.157)key figure | The central comparison: slope, R² and RMSEE for the area method and PLS at 25, 50 and 75% PVP. | F2, K2 | RMSEE is a calibration error, not a validated prediction error. |
| Table 2 (p.158)key figure | R2Y of each PLS component: how much each component explains at each PVP level. | F3, K4 | Cumulative R2Y differs from the Table 1 PLS R². |
| Fig. 3 (p.158)key figure | Loading weights of components 1 and 2 at 25% PVP, with the crystalline felodipine pattern.· | F3, K4 | No zero line; signs not readable. |
| Fig. 4 (p.158) | Loading weights of components 1 and 2 at 50% PVP.· | F3 | No zero line; signs not readable. |
| Fig. 5 (p.158)key figure | Loading weights of components 1 and 2 at 75% PVP.· | F3 | No zero line; signs not readable. |
| Fig. 6 (p.159)key figure | PLS predictions against actual crystallinity with prediction bands at three PVP levels.· | F2, K2 | Calibration samples only; no test set. |